Title:
 
Measurement of Flash Solar Simulator Output Spectra over Bulb Lifetime and the Effects on Spectral Mismatch
 
Author(s):
 
M. Bliss, A. Smith, B. Mihaylov, M. Krawczynski, T.R. Betts, R. Gottschalg
 
Keywords:
 
Spectroradiometer, Solar Simulator, Spectrum, Uncertainty, Spectral Mismatch
 
Topic:
 
COMPONENTS FOR PV SYSTEMS
Subtopic: Standardisation as Tool for Innovation and Cost Reduction
Event: 28th European Photovoltaic Solar Energy Conference and Exhibition
Session: 4AV.6.46
 
Pages:
 
3505 - 3509
ISBN: 3-936338-33-7
Paper DOI: 10.4229/28thEUPVSEC2013-4AV.6.46
 
Price:
 
 
0,00 EUR
 
Document(s): paper, poster
 

Abstract/Summary:


The effects of solar simulator xenon flash lamp ageing and spectral output measurement uncertainty on the spectral mismatch are investigated. Initial measurements demonstrated that the spectrum of an older lamp set has a relatively lower blue and larger red content compared to a new set of bulbs. Measurements over the life-time of several sets of bulbs however, showed large unexpected variations due to measurement uncertainty in the spectral measurements themselves. The main influencing factors are investigated and a faulty temperature control is found to be the main source of uncertainty. It is shown that this alone can affect the mismatch calculation to a larger degree than the MMF would correct in itself.